| Current Range: | 2 nA - 2 mA |
| Resolution: | 100 fA |
| Input Voltage Burden: | < 200 µV on all current ranges |
| Display: | Digital panel meter with 4 ½-digit resolution |
| Display Reading Rate: | 3/s |
| Ranging: | Manual or remote controlled (optional) |
| Overrange Indication: | Display flashes "0000" |
| Current Source Bias: | Push-button selectable: -90V, 0V, +90 V |
| Analog Output: | Scaled voltage output, 2V full scale on all ranges |
| Zero Adjustment: | Through two potentiometers on rear panel |
| Common Mode Voltage: | Maximum 90 V |
| Input Connector: | BNC on front panel |
| Input Protection: | Overvoltage and arc protected current input |
| Isolation: | Meter common to chassis: typically > 100 GOhm |
| Dimensions: | 130 mm high x 110 mm wide x 170 mm deep |
| Weight: | 1.5 kg |
| Power: | 100-240V, 50-60Hz, 10VA |
| Remote Connector: | 9-pin Dsub connector on rear panel |
| Input Signals: | 3 TTL-signals for range selection |
| 1 TTL-signal for selecting local or remote control | |
| Output Signals: | 3 TTL-signals for reading the selected range |
| 1 analog signal for current monitoring (0-2V or 0-10V) |
|
Range
|
Max
|
Display
|
Accuracy
(1 day)
|
Temp
Coefficient
|
Analog
|
|
.
|
Range
|
Resolution
|
+-(%Reading
+ %Range) @ 25°C
|
+-(%Reading
+ %Range)/°C
|
Rise
Time
|
|
pA
|
1999.9
pA
|
0.1pA
|
0.5
% + 0.05%
|
0.02% + 0.002%
|
100
ms
|
|
nA
|
199.99
nA
|
10
pA
|
0.1
% + 0.05%
|
0.002%
+ 0.002%
|
100
ms
|
|
uA
|
19.999
uA
|
1
nA
|
0.1
% + 0.05%
|
0.002%
+ 0.002%
|
10
ms
|
|
mA
|
1.9999
mA
|
100
nA
|
0.1
% + 0.05%
|
0.002%
+ 0.002%
|
1
ms
|
| Digital Panel Meter Accuracy: +/- 1 digit , temperature drift: 0.4 counts/°C | |||||
| Analog Output Accuracy: 2V full scale on all ranges, accuracy 1% of reading + 2mV offset | |||||
|
C4
Scientific Systems
754 Leona Ln Mountain View, California 94040 Phone: (650) 961-6620 Fax: (650) 961-8591 support@c4scientific.com |